The 5700A and 5720A are five-function calibrators designed to address the most demanding electrical calibration workload. They source direct and alternating voltage and current and resistance. A wideband voltage option extends overall ac bandwidth to 10 Hz to 30 MHz to cover to RF voltmeters. Both are compatible with the 5725A, and 5205A/5215A amplifiers.
Fluke 5700A Features:
| Voltage DC |
| Ranges: |
0 to ±1100V |
| 5700A Best 1 Year Specification: |
±7.0 ppm of setting |
| 5720A Best 1 Year Specification: |
±3.5 ppm of setting |
|
| Voltage AC |
| Ranges: |
220 mV to 1100V |
|
10 Hz to 1 MHz |
| 5700A Best 1 Year Specification: |
±75.0 ppm of setting |
| 5720A Best 1 Year Specification: |
±45.0 ppm of setting |
|
| Resistance |
| Ranges: |
0 to 100 MO, 18 values in x1 and x1.9 |
| 5700A Best 1 Year Specification: |
±12.0 ppm |
| 5720A Best 1 Year Specification: |
±8.5 ppm |
|
| Current DC |
| Ranges: |
0 to ±2.2A (11A with 5725A) |
| 5700A Best 1 Year Specification: |
±50 ppm of setting |
| 5720A Best 1 Year Specification: |
±35 ppm of setting |
|
| Current AC |
| Ranges: |
9 m A to 2.2A (11A with 5725A) 10 Hz to 10 kHz |
| 5700A Best 1 Year Specification: |
±140 ppm of setting |
| 5720A Best 1 Year Specification: |
±120 ppm of setting |
|
| Wideband AC Voltage Option |
| Ranges: |
300 m V to 3.5V |
|
10 Hz to 30 MHz |
| 5700A Best 1 Year Specification: |
±0.4% of setting |
| 5720A Best 1 Year Specification: |
±0.4% of setting |
|
| General Specifications |
| Warm up Time |
2x the time since last warmed up, to a maximum of 30 minutes |
| System Installation |
Rear output configuration and rack-mount kit available |
| Interfaces |
IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in (BNC), phase reference out (BNC) |
| Temperature Performance |
Operating: 0°C to 50°C
Calibration: 15°C to 35°C
Storage: -40°C to 75°C |
| Relative Humidity |
Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage: < 95%, non-condensing. A power-on stabilization period of four days may be required after extended storage at high temperature and humidity.
|
| Safety |
Designed to comply with UL311; IEC 348-1978; IEC 66E (CO)4; CSA 556B |
| Guard isolation |
20 volts |
| EMI/RFI |
Designed to comply with FCC Rules Part 15, Subpart J, Class B; VDE 0871, Class B |
| Line Power |
47 to 63 Hz; + 10% allowed about selectable nominal line voltage: 100V, 110V, 115V, 120V, 200V, 220V, 230V, 240V --------------
Maximum power: 5700A/5720A: 300 VA;5725A, 750 VA
|
| Size |
5700A/5720A: Height 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet; width 43.2 cm (17 in), standard rack width; depth 63.0 cm (24.8 in) overall; 57.8 cm (22.7 in), rack depth
5725A:Height 13.3 cm (5.25 in); width and depth same as 5700A/5720A
Both units project 5.2 cm, (2 in) from rack front |
| Weight |
5700A/5720A:27 kg (62 lbs.)
5725A: 32 kg (70 lbs.) |
Fluke 5700 A Standard Accessories
The following items are shipped by Global Test Equipment with each unit ordered.
| Part Number |
Description |
Qty |
| 601622 |
Operator's Manual |
1 |
| 1668111 |
Getting Started Guide |
1 |
| 601655 |
Remote Programming Reference Guide |
1 |
| 601630 |
Service Guide |
1 |
| |
Service Manual Supplement 8/03 |
1 |
| 601648 |
Operator's Guide |
1 |
| |
Power Cord |
1 |
OPTIONS***
| Option |
description |
| 03 |
Wideband AC Voltage Module |
*** The above list of options represents commonly found options. Units available from Global Test Equipment may have a combination of these options. Contact Us for detailed information about the options available on the units in stock.
|